ING Banner
Home > Astronomy > ISIS > Image Polarimetry


Image Polarimetry with ISIS

  1. Configuring the spectrograph for image polarimetry
  2. Configuring the telescope
  3. Aquaring objects and taking data
  4. Some useful numbers

1. Configuring the spectrograph for image polarimetry

ISIS is designed to work as a spectrograph, but by replacing the grating for a flat mirror and by taking away the spectrograph entrance slit we create an imaging system. With some additional optical components ISIS can be made into an imaging polarimeter by making use of the existing waveplates that are an integral part of ISIS. To facilitate imaging polarimetric observations, apart from changing the grating for a mirror, a wide calcite analyzer needs to be inserted, together a Dekker comb mask, and a passband filter. Finally, focus adjustments need to be made. This documents describes in detail how this is accomplished. It requires several modifications of the setup that have to be done by the SA early in the afternoon.

1.1- Mount the Savart (calcite) analyzer (see Fig. 1) in the multislit position of the ISIS slit carriage.  You may have to remove the image slicer. This operation should only be done by the support astronomer.

First put the dichroic out of the beam for easy access:

SYS@taurus> bfold 0

Then protect the slit unit with the dekker:

SYS@taurus> dekker 1

Then move the multislit unit to the light path

SYS@taurus> mslit

Then unlock the slit door:

SYS@taurus> slit_door open

You can open the slit door located on top of the red cryostat. Inside you'll have access to the the slit unit, dicroic, filters and dekker. Carefully slide out the image slicer (or whatever unit in the multislit unit) and put it in its box and  store it in the WHT cabinet.  Take the analyzer unit out of it's  box and carefully slide it in and lock it's pin holder.

If you want to change filters (see next section) proceed in a similar way with the filter units, just pull them out, put the units back to its box and introduce the other ones in the unit. Be careful to arrive to the end and that the filter units are locked.

Close the slit door by hitting it's lockers. Finally lock the slit door in the control window:

SYS@taurus> slit_door close


calcite unit
Fig. 1 -  The analyzer unit for image polarimetry. The unit should be mounted in the multislit mask unit of
ISIS. The calcite is below the field mask (in yellow).

1.2 - Mount the mirror at the appropriate position in place of the grating. There is a flat mirror mounted in a ratting mount. Just replace the ratting with the flat mirror unit in the same way as changing grating. This operation should only be done by the support astronomer or telescope operator (if trained).

SYS@taurus> setgrating red R1200R

to change the red grating to the mirror one, any valid name of a ratting works fine with the mirror as it will be used in order 0. The same for the blue arm.
Then answer yes to access the grating door. The doors are then unlocked and can be open manually. To release the gratings use the open/close buttons. Lock the grating door by hitting the lockers. Back at the console set the central wavelength to 0.

SYS@taurus> cenwave red  0

The same if you plan to use the blue arm.

Notice that there is only one flat mirror unit, so imaging polarimetry can only be done with one arm at a time.


1.3- Mount the filters (colour and/or neutral density) in the modified ISIS below slit filter slides (see Fig. 2). This operation should only be done by the support astronomer. The filter slides are then mounted in their position below the ISIS slit unit following the same procedure used to mount the analyzer unit (in fact is a good idea to change the analyzer and filters together).  Only two filters can be mounted on both RED and BLUE arm, but remember that only one arm can be used.


filter unit

filter

Fig. 2 - Upper image is one of the two filter units of the RED arm. Other two can be mounted in the blue
arm. Lower image is the R filter mounted in the filter unit. Notice that black tape at the two sides of the
5x5cm square filters are needed to avoid stray light.


Normal filter slides holds two filters, while the modified one holds only one. Due to the positions of the filters in the modified slides, it vignettes the beam in any of the two positions of the normal filter unit. To overcome this problem and minimize vignetting, it is necessary to insert the filter slide to a position 4 which is between normal positions 2 and 3 (in position 1 filter slide is OUT). To use a filter mounted in BFILTB slide type

SYS@taurus> bfiltb 4

You can use filters in the following slides: RFILTA, RFILTB, BFILTA, BFILTB.

1.4 - Put the half-wave plate in the beam in preparation for observing by typing HWIN (HWOUT takes it out of the beam again).

SYS@taurus> hwin

1.5 - Setup the CCD by setting an appropriate window and readout speed, as advised by your support astronomer

SYS@taurus> window red 1 "[730:2148,1350:2150]"

SYS@taurus> window blue 1 "[700:2148,1600:2500]"

1.6 - Setup ISIS to take a lamp exposure

SYS@taurus> agcomp

SYS@taurus> complamps w

SYS@taurus> compnd 2 (or whatever neutral density filter is required).

This setup produced flats like the one presented in Fig. 3

flat

Fig. 3 - This is the smaller window that include all the field and the bias columns of the CCD (left end of the image).
The double pattern of slits correspond to the ordinary and extra-ordinary image of the comb mask as produced by
the calcite block.

1.7 - Focus the spectrograph by moving the collimator. First set the collimator to a rough focus using the commands

SYS@taurus> rcoll 10000   (for the red arm)

SYS@taurus> bcoll 5500   (for the blue arm)

Then step in units of 500 collimator units about this nominal focus taking a GLANCE exposure at each setting until the edges of the dekker are at their sharpest (as determined e.g. with imexam). The spectrograph will now be in focus.

1.8 - Check the polarization optics by putting a polaroid in the beam using the

SYS@taurus> mainfiltc MF-POL-PAR

Now you should check zero angles of the half-wave plate by taking images around expected zero angle and looking at the difference in the intensity between the ordinary and extraordinary beams. The zero angles measured on August 3 2012 for the blue arm were 6, 51, 28.5 and 73.5 degrees instead of 0, 45, 22.5 and 67.5 degrees. At zero angles around 6 and 51 degrees, the intensity difference between the ordinary and extraordinary beam should be maximal, and at angles around 28.5 and 73.5 degrees the intensity difference should be minimal.

These angles must be introduced in the imaging polarimetry observer's script at /home/whtobs/impolscript. Also, make sure that the camera name is updated in the script.

Remember to remove the polaroid with the MAINFILTC OUT command.

Back to the top

2. Configuring the telescope

2.1 - Determine the rotator centre on the direct-view mirror by typing AGCOMP and asking the telescope operator to determine the rotator centre on this mirror.  Put a marker on the DS9 display. It is also a good idea to write down the TV coordinates of this point in case the marks are erased.

2.2 - Perform a calibrate about this point on the direct-view mirror (the telescope operator will do this for you).
  
2.3 - Crudely refocus the telescope by subtracting 0.60 mm from its ISIS slit value (typically 97.95 mm) using the ICL command FOCUS 97.35 (or whatever). This will enable you to take exposures and view the star roughly in focus on the CCD and on the slit-viewing mirror. To ensure that the star is also in focus on the direct-view mirror, set TVFOCUS 15000 in TVSCALE 5 at ICL.
  
2.4 - Determine the position of the centre of the central dekker on the direct-view mirror:
- switch to AGSLIT mode, point the telescope to a bright star and take a glance image with ISIS.
- identify the star in the image, ask the TO to apply offsets until the star appear centered on the central dekker
- switch to direct view using AGCOMP and take an image with the direct view camera.
- Mark the position of the star on the direct-view display and note the coordinates of the point in case the marks are erased.  This position will be used to identify and roughly center the targets with the direct view camera.

2.5 - To center the targets just use the direct view mode, identify the target and the with the help of the TO, center it in the marker done in 2.4. Then remove the direct view mirror using AGSLIT, and take a glance image in ISIS. Then do a final tuning by asking the TO to introduce the offsets to center the target in the dekker mask.
  
2.6 - Refocus the telescope accurately using the usal focus procedure. First, setup the spectrograph for observing on-sky by typing AGSLIT and ROT SKY 0. Then acquire an (of order) 11th magnitude star and centre it in the centre of the central slot of the dekker mask as found in the previous step. Take a test exposure using GLANCE to determine the optimum exposure time and use FOCUSRUN command in the usual way.
  
2.6 - Refocus the autoguider by subtracting 2000 from its ISIS slit value using the ICL command AUTOFOCUS 2000 (or whatever). You are now ready to observe.


Back to the top

3. Acquiring objects and taking data


3.1 - Acquire objects by typing

SYS@taurus> agcomp

and positioning the object of interest on the desired slot, as determined by the marks drawn on the TV screen.
Then type

SYS@taurus> agslit

and the object should now fall directly onto the desired dekker slot. Hence this acquisition procedure does not require one to move between LSLIT and MSLIT, which would compromise the flat-fielding (since the positions of the dekker mask edges as projected onto the detector would continuously be changing).
  
3.2 - Update a target name and exposure time in /home/whtobs/impolscript and take data by typing

SYS@taurus> source impolscript

This will take 4 images at zero angles determined earlier. Fig. 4 is an example of an image of a 0 polarization standard.

image

Fig. 4 - image of a 0 polarization standard star.

Back to the top

4. Some useful numbers


4.1 - There are 7 slots in the dekker mask and hence 14 slot images on the CCD (due to the action of the calcite block).
 
4.2 - Each slot is approximately 6 arcseconds wide and the throw of the calcite is approximately 7 arcseconds. Hence there is a gap of width 1 arcsecond between the o and e rays. The masked spaces between the slots are approximately 11 arcseconds in width.
  
4.3 - The image of the calcite block obtained on the CCD is approximately 132 by 111 arcseconds in size. Each pixel is approximately 0.22 by 0.22 arcseconds in size with the REDPLUS, and 0.2 by 0.2 arcseconds with the EEV12.
  
4.4 - If the rotator position angle on the sky is zero, the slots of the dekker are aligned approximately east-west.


Top | Back

Contact:  (ISIS Instrument Specialist)
Last modified: 14 August 2012